Shop/ Sand Scientist Series/ Fully Automatic Loss on Ignition Tester VLOI-FA
Sand Scientist Series Fully Automatic Loss on Ignition Tester VLOI-FA
Sand Scientist Series · Equipment for Green Sand · Testing Equipment - Green Sand · Equipment for Resin Coated Sand (Hot) · Testing Equipment - Resin Coated Sand

Fully Automatic Loss on Ignition Tester VLOI-FA

The Versatile VLOI-FA automates the loss on ignition test: it weighs the sample, fires it at 982 °C, cools it and weighs it again, reporting the result to AFS 5100-12-S with a record kept for the quality file.

Green sandResin coated

Prefer email? Write to [email protected] and mention this model.

Since 1966. Designed and built by Versatile, trusted by 9,000+ foundry customers.
Specifications

Technical specifications

Test
Loss on ignition of foundry sand, bentonite and pre-blends
Method
AFS 5100-12-S
Firing temperature
982 °C (1800 °F), with optional drying at 107 °C
Measuring range
LOI 0 to 5 per cent, depending on the application
Balance
Integrated, ±0.005 g
Sample mass
25 ±5 g sand; 8 ±3 g bentonite or pre-blends
Sand types
Green sand, resin coated, no-bake, cold box and CO2 sand
Data
Wired network connection, optional Wi-Fi, V-Sync cloud sync
Dimensions (W x D x H)
780 x 760 x 410 mm
Weight
Approximately 26 kg
Traceability
Traceable to national standards, NABL-traceable calibration at least once a year
Power supply
230 V AC, 50 Hz standard; other supplies such as 110 V, 60 Hz on request. 15 A at 230 V
Overview

See it explained: the mass loss curve, why the run ends on burn rate rather than a timer, and holding LOI on a control chart, on our Loss on Ignition feature page.

The Versatile VLOI-FA runs the loss on ignition test from start to finish. It weighs the sample, fires it, cools it and weighs it again, then reports the loss as a percentage to AFS 5100-12-S. The firing step completes on a burn rate check rather than a fixed timer, so the result follows the sample instead of the clock.

Application

Loss on ignition measures what burns away from a sand sample. In green sand it tracks the carbon carriers such as sea coal, along with the residue that builds up in a running system. In resin coated and chemically bonded sand it indicates how much binder sits on the grain. Both readings feed decisions that show up in the casting: gas defects and veining when the figure runs high, weak bonding and poor surface when it runs low. The test also applies to bentonite and pre-blends.

Supplied with: the main unit with integrated furnace and balance, three refractory crucibles and tongs. A Class F1 or E2 calibration weight set and a temperature verification kit are recommended alongside, and the unit needs adequate ventilation with an exhaust.

Safety & care. The furnace runs hot. Let it cool before handling crucibles, use the tongs supplied, and keep the exhaust clear.
Connected instrument

Readings, in the cloud with VSync

This model works with VSync web sync, so results are logged and stored online. A supervisor can review a shift from a browser, and a service engineer can look at the same data remotely. See how VSync works →